Almost every red light panel on the market is sold with an irradiance figure in mW/cm². Very few are sold with any account of where that figure came from. Where a brand does offer a laboratory certificate, these are the parts of it that actually carry weight — and the parts that are easiest to leave out.

The accreditation, not the laboratory's name

Any company can call itself a testing laboratory. An ISO/IEC 17025 accreditation is different: it is granted by a national accreditation body, it is periodically re-audited, and it comes with a registration number that can be looked up independently.

Worked example. The Dermfix SLF1800 was measured by LightingLab Calibration Laboratory Ltd. of Veszprém, Hungary, under accreditation registration NAH-2-0368/2023. It is that registration number, rather than the laboratory's name, that lets anyone confirm the accreditation is real and current.

A certificate reference you can quote back

A measurement that actually happened produces a document, and that document has an identifier. For the SLF1800 it is certificate LL-26-6174, issued 2 November 2026.

An irradiance claim with no reference number attached cannot be traced to anything. It may still be accurate. There is simply no way to find out.

Measurement uncertainty

This is the detail that separates metrology from marketing, and it is almost never quoted.

Every physical measurement has a margin of error, and a laboratory report states it. The SLF1800 certificate records an expanded uncertainty of ±2.3% at a coverage factor of k=2 — meaning roughly a 95% confidence interval. A brand quoting a bare number to one decimal place, with no uncertainty and no coverage factor, is quoting a reading off a screen rather than a measurement.

Cold, or at working temperature

LEDs lose output as they warm up. A panel measured the instant it is switched on will always read higher than the same panel ten minutes into a session — and ten minutes into a session is when you are actually using it.

The SLF1800 was recorded both ways: 83.3 mW/cm² at switch-on and 74.0 mW/cm² after ten minutes, both at 15 cm. Where only one figure is published, it is almost certainly the first.

Centre of the panel, or averaged across it

Light is brightest in the middle of a panel and falls away towards the edges. This is physics, not a defect — but it means a centre-point reading and an area average describe two different things.

Neither is dishonest. They are simply not interchangeable, in either direction. If one brand publishes a centre figure and another publishes an area average, comparing the two numbers tells you very little. Check which you are looking at before drawing any conclusion.

The distance, stated

An irradiance figure without a measuring distance is meaningless. 15 cm has become the industry convention, though nothing requires it. The SLF1800 figures above are at 15 cm; the same panel measures 73.8 mW/cm² at switch-on and 65.7 after ten minutes at 30 cm.

A short checklist

  • Is the accreditation body and registration number given, not just a laboratory name?
  • Is there a certificate or report reference?
  • Is a measurement uncertainty stated, with a coverage factor?
  • Was the panel measured cold or at working temperature — and does the brand say which?
  • Is the figure a centre reading or an average across the panel face?
  • Is the measuring distance stated?

Six questions. Most published irradiance claims fail at the first one.